What It Is

The result: every element originally locked in the solid matrix ends up as free ions in a clear solution, ready for measurement by ICP-OES, ICP-MS, atomic absorption, or other elemental techniques.

Why It Exists

The bottleneck

A modern ICP-MS takes under two minutes per sample; preparation takes hours. The expensive spectrometer sits idle while preparation catches up.

Where the errors are

Incomplete digestion means low, imprecise results; volatile loss under-recovers mercury and arsenic; contamination from open vessels inflates results.

Where the hazards are

Concentrated hot acids, high pressures, potentially explosive chemistry in open beakers.

The Science

Why microwave works better.

HotplateHot BlockMicrowave (closed vessel)
Time2–8 hrs1–4 hrs20–45 min (batch) / ~5 min (sequential)
Max temperature~120 °C~150 °C200–260 °C
Volatile retentionPoorPoorExcellent
Contamination riskHighModerateLow
Complete digestion of refractory matricesOften notOften notAchieved
Automatic documentationNoNoYes
Batch vs Sequential — The Core Decision

Batch (MARS 6, MARSXpress 2.0)

All samples processed simultaneously; one method applies to all; highest throughput for similar matrices; best for high-volume homogeneous sample types.

Sequential (BLADE)

Each sample processed individually under its own method; mixed matrices possible in the same run; approximately 5 minutes per sample; best for varied, mixed or daily-intake laboratories.

Acid Selection Guide
MatrixAcid combinationNotes
Most organics, metals, soilsHNO₃ alone or with H₂O₂Oxidises organic matter, dissolves most metals
Precious metals (Au, Pt, Pd)HNO₃ + HCl (aqua regia)The only combination dissolving gold
Silicates (rock, soil, glass, ceramic)HNO₃ + HFHF essential to attack Si-O bond
Refractory organics, high-lipidHNO₃ + H₂O₂H₂O₂ boosts oxidising power
High-temperature workHNO₃ + H₂SO₄H₂SO₄ raises boiling point

HF requires special vessel materials and handling precautions; always refer to the specific vessel compatibility guide.

Standard Methods Using Microwave Digestion

USEPA 3051A (sediments, sludges, soils, oils) · USEPA 3052 (siliceous and organically based matrices, with HF) · USEPA 3015A (aqueous samples) · EN ISO 54321 (soil, biowaste, sludge, waste) · EN 13805 (foodstuffs) · ICH Q3D / USP <233> (pharmaceutical elemental impurities) · AOAC Official Methods (food and feed).